IR-2: Fourier Transform Infrared (FTIR) Spectrometer


Model: Digilabs FTS-6000
Location: 104 Partners III
Methodology and Rate: Vibrational (VIB) -
Current Instrument Status: Operational
Training
Operating Procedures
- General IR-2 operating procedures
- Methodology specific operating procedures
Helpful Links
IR-2 workstation is a Digilabs FTS-6000 is a step scanning FTIR that is
configured for the Mid-Infrared region only. It is also the infrared source
for IR-3 the UMA-500 Infrared Microscope. Measurements can be made
in transmission, reflection, and attenuated total reflection (ATR) with this
instrument. Methodologies that are most often conducted with this instrument
are routine sample measurement, micro-sample measurements, fast TRS (Time
Resolved Spectroscopy), cryogenic work and laser photolysis. The IR-2
workstation can also be fitted with a wide range of FTIR accessories to
provide additional capabilities.
Applications
- Spectrum measurements (transmission, absorbance, reflection) over a
wavenumber range of 7,500–450 cm1 (1.3μm–22.0μm)
- Time-resolved spectroscopy (TRS) for fast kinetic process studies (200ns)
- Phase-resolved spectroscopy (PRS) for modulation experiments
- Self assembled monolayer (SAM) characterization
- Synthesis characterization
- Reaction monitoring
- Materials analysis
- Defect and surface contamination analysis
- Laboratory instruction in spectroscopy
- Spectroscopic research experiments
- Laser photolysis
- Cryogenic characterization of samples
- Attenuated Total Reflection (ATR)
- Diffuse Reflection (DR)
- Grazing Angle (GR)
Configurations
- Beam Splitter
- Extended range KBr (Mid-Infrared)
- Detectors
- Liquid Nitrogen cooled MCT-A (Narrow band 7,500–650 cm-1)
- Liquid Nitrogen cooled MCT-B (Wide band 7,500–450 cm-1)
- Source
- Water-cooled SiC Ceramic (Mid-Infrared)
- Additional
- The attached UMA-500 Infrared Microscope
- The spectrometer and sample compartment are purged with dry compressed air,
which reduces the possibility of atmospheric water or CO2 contamination of the spectra and samples.
- A wide range of sample compartment FTIR accessories are available
- “Win IR-PRO” software for complete data collection and processing
Specifications
- Spectral Range: 7,500-450cm-1
- Maximum Resolution: 0.1 cm-1
Note: Laser photolysis requires additional training. Please see Training for more information